10x Running Buffer for 3130, 3730, 3500/SeqStudio™ Flex

  • 10x Running Buffer for 3130, 3730, 3500/SeqStudio™ Flex
10x Running Buffer for 3130, 3730, 3500/SeqStudio™ Flex

10x Running Buffer for 3130, 3730, 3500/SeqStudio™ Flex

10X running Buffer


Catalog Number SG6121X


Product description

10X running Buffer is mainly applied to the electrophoresis buffer of sequencing instruments, which can provide an electrophoresis buffer system for sequencing process after dilution to 1X. This product has the advantages of strong buffer capacity and low background of sequencing results. After testing, 20 consecutive plates still have good buffer capacity.

Contents

Item No.

Contents

specification

Description

SG6121X

10X running Buffer

500mL

Electrophoretic buffer for sequencing machines

Use steps

• The 10X running Buffer and double steaming water are configured with 1X working liquid 150ml according to the volume ratio of 1:9

• Remove the anode buffer bottle and cathode buffer tank from the sequencing instrument and clean them with double steaming water

Pour 1X working solution into anode bottle and cathode buffer tank to mark.

Assemble the anode bottle and cathode buffer tank into the sequencer

 

Problem and solution

If 1X working solution configured with 10X running Buffer has a great impact on the sequencing results, the following phenomena usually occur:

Phenomenon

Reason

Solution

The current line is unstable during electrophoresis

The ionic strength in the buffer is not enough

Replace with a new 1X working fluid

It has a blue or red background

The buffer is mixed with organic matter

Clean the buffer bottle and tank, and check the plastic pipe connecting the pump to the anode bottle

The current is too high during pre-electrophoresis

(greater than 1200mA)

The ionic strength of buffer is too high

Reconfigure 1X working fluid

Peak delay or backward

The buffer is mixed with surfactant

Reconfigure 1X working fluid

Storage conditions

The temperature of the storage environment (refrigerator) should be 4 - 8 °C, too low temperature will make the reagent freeze, in the case of incomplete melting will lead to the ionic strength of the reagent is too high, so when configuring the working liquid, pay attention to the temperature of the refrigerator and whether there is ice in the bottle, if found, it should be completely dissolved and mixed before use.