POP-4 Polymer
Catalog Number: SG5123X
For use with 3500⁄3500xl series instruments
WARNING! Read the Safety Data Sheets and follow the handling instructions. Wear appropriate protective eyewear, clothing, and gloves.
The polymer is the separation matrix for capillary electrophoresis. It is supplied as a ready-to-use pouch with a radio frequency identification (RFID) tag incorporated into the label. The instrument uses the RFID tag to track polymer usage and expiration.
|
Item No. |
Amount |
Storage |
|
SG5123X-28 |
1 x 28mL bottle [1] |
Store at 2–8°C |
|
SG5123X-7 |
1 x 7mL bottle [1] |
[1] The polymer has been validated for HID applications.
|
Pouch size |
Instrument |
On-instrument supported limits[1] Lower of: |
Guidelines |
|
96 samples |
8-capillary |
14 days, 96 samples, 12 injections, or expiry date |
The polymer has been verified for use for up to 14 days on the instrument. |
|
24-capillary |
14 days, 96 samples, 5 injections, or expiry date |
||
|
384 samples |
8-capillary |
14 days, 384 samples, 60 injections, or expiry date |
|
|
24-capillary |
14 days, 384 samples, 20 injections, or expiry date |
||
|
960 samples |
8-capillary |
14 days, 960 samples, 120 injections, or expiry date |
|
|
24-capillary |
14 days, 960 samples, 50 injections, or expiry date |
[1] The pouch has adequate polymer to support the stated number of samples or injections, plus additional volume to accommodate installation and wizard operations. Multiple pouch installations and/or excessive use of wizards reduce the number of remaining samples and injections. For example, if you run the total bubble remove option in the Remove Bubbles wizard more than four times, the number of remaining samples and injections is reduced.
Replenish polymer or change polymer type
1. Check the expiration date on the label to ensure that the polymer is not expired and will not expire during intended use.
IMPORTANT! Do not use if the product is expired, if the pouch or label is damaged, or if the top seal is missing or damaged.
2. Allow the refrigerated polymer to equilibrate to room temperature (15–30°C) before use.
3. In the instrument software, go to the Wizards menu, then click Replenish Polymer (requires 10–20 minutes) or Change Polymer Type (requires 60–70 minutes).
Note: If you are changing the polymer type, you also need Conditioning Reagent.
4. Follow the prompts in the Wizard window.
5. When instructed to install the polymer, peel off the seal at the top of the pouch fitting.
Note: You may notice a tiny droplet of polymer inside the fitting (residual from the pouch filling process). This is not expected to cause any performance issues.
To install the polymer on the instrument and start the run, see your instrument user guide.
Important notice regarding use of consumables that exceed supported limits
BEFORE DISMISSING THE WARNING THAT THE CONSUMABLES HAVE REACHED SUPPORTED LIMITS AND CONTINUING WITH OPERATION OF THE INSTRUMENT, PLEASE READ AND UNDERSTAND THE FOLLOWING IMPORTANT NOTICE AND INFORMATION:
Na Yuan Tech. does not recommend the use of consumables that exceed supported limits. The recommended limits are designed to promote the production of high quality data and minimize instrument downtime. Reagent and consumable lifetime minimum performance are based on testing and studies that use reagents and consumables that have not exceeded supported limits. The use of consumables beyond the supported limits may impact data quality or cause damage to the instrument or capillary array. The cost of repairing such damage is NOT covered by any our product warranty or service plan. Customer use of expired consumables is at customer's own risk and without recourse to our company.